What are IEC 60601-compliant connectors?
Meet the Connector: IEC 60601-Compliant Connectors
IEC 60601 is a technical standard that ensures the basic safety and essential performance of electrical medical equipment and systems for use in the home healthcare environment. IEC 60601-compliant connectors adhere to a specific pin layout, meet insulation requirements, feature specific grounding measures, as well as protect against electrical hazards to patients and medical staff. IEC 60601-1 is the third edition of this standard. It is reviewed every five years and will undergo review in 2025.
As essential components that are part of the medical device, IEC 60601-1 compliant connectors regulate the interfaces that enable smooth communication between different components. Medical devices are usually connected to the general power supply network and thus pose potential danger to patients and operators. IEC 60601-1 defines clear requirements to ensure that such devices function reliably under the intended conditions of use while minimizing potential hazards.

Winchester Interconnect offers a range of IEC 60601-compliant interconnect products such as nano-miniature connectors that can serve in patient monitoring devices and other high-reliability applications.
Four core requirements apply to interconnects:
- Electrical safety and isolation (IEC 60601-1): Creepage and clearance distances, leakage current limits, touch-proof design, and IP rating
- Electromagnetic compatibility (IEC 60601-1-2): Shielding and EMI protection and EMI filters
- Mechanical requirements (IEC 60601-1 & IEC 60601-1-11 for home use devices): Durability and mating cycles, misconnection prevention, and mechanical strength
- Usability and human factors (IEC 60601-1-6 & IEC 62366): Ease of Use
Patient considerations: MOPP and MOOP
The 60601-1 standard specifies means of patient protection (MOPP) and means of operator protection (MOOP) for devices that have contact with a patient or operator. These protection ratings specify stringent creepage and clearance distances. (Clearance is the shortest distance through air between two conductive parts, such as pins. Creepage is the shortest distance between two pins along the insulating material between them.) These distances are meant to ensure that no arcing can occur between pins, causing potential harm to patients and device handlers.
The standard also specifies a maximum leakage current that flows from the probe contact to ground through the patient. The latest revision increased protections by raising the ESD protection threshold from 8 kV to 15 kV; increasing MOPP/MOOP values and re-definition of clearance and creepage requirements.
Most commercial interconnects cannot meet the strict requirements of 60601-1. Connector suppliers that offer compliant components work closely with regulatory bodies to design products that meet the highest levels of compliance for safety, performance, and reliability to prevent risks such as electrical shock, misconnection, or electromagnetic interference.

Schurter’s EC12 filter series is available as a standard or medical M5/M80 version and is used in devices that are especially sensitive to interference. This includes IT or telecom systems according to IEC 92368-1 and medical equipment according to IEC 60601-1. Filter versions with increased surge withstand voltage capability are also available.
Design Notes
Standardization: The first edition of IEC 60601 was published in 2010. The current version of the standard, IEC 60601-1-11:2015, was issued in 2015 and amended in 2020. This second edition cancels and replaces the first edition of IEC 60601-1-11, published in 2010, and constitutes a technical revision. It is subject to review every five years and is up for review in 2025.
The most significant changes in the latest edition include:
– Correction of test method for relative humidity control at temperatures above 35 °C
– Redrafting of subclauses that altered instead of adding to the general standard or other collateral standards
– Harmonizing with the changes to the amendments to the general standard and other collateral standards
IEC 60601 has collateral (IEC 60601-1-X) and particular standards (IEC 60601-2-X) for specific device types, such as:
- IEC 60601-2-24 for infusion pumps
- IEC 60601-2-22 for laser equipment
- IEC 60601-2-37 for ultrasound diagnostic systems
By specifying products that meet 60601-1 requirements, designers can simplify the associated necessary risk analysis.

ODU MINI-SNAP connectors, available from Mouser Electronics, fully comply with the requirements of IEC 60601-1. These mainstays of medical environments are extremely durable and meet the requirements for cleaning and autoclavability.
Markets, Sectors, and Applications
Medical, Test & Measurement
The 60601-1 standard applies to a wide range of medical devices. Three classifications address cardiac contact: Type B devices have no contact with a patient’s heart (hospital beds, surgical lights, infusion pumps). Type BF devices have direct patient contact but not heart contact (endoscopes, ultrasound equipment, transesophageal echocardiogram, probes). Type CF devices have direct contact with the patient’s heart (heart monitors, electrocardiogram machines).
Suppliers
EDAC, JAE, LEMO (PEI Genesis), SCHURTER, TE Connectivity, Winchester Interconnect, ODU (Mouser Electronics)
Related products:
Like this article? Check out our other Meet the Connector and Connector Basics articles, our Medical Market Page, and our 2024 and 2025 Article Archives.
Subscribe to our weekly e-newsletters, follow us on LinkedIn, Twitter, and Facebook, and check out our eBook archives for more applicable, expert-informed connectivity content.
- What is Semi-Rigid Coax Cable? - March 11, 2025
- Meet the CEO: Weidmuller USA’s Randy Sadler - March 4, 2025
- What are MIL-DTL-22992? - February 25, 2025